The Number of Microbubbles Generated During Cardiopulmonary Bypass Can Be Estimated Using Machine Learning From Suction Flow Rate, Venous Reservoir Level, Perfusion Flow Rate, Hematocrit Level, and Blood Temperature

IEEE OPEN JOURNAL OF ENGINEERING IN MEDICINE AND BIOLOGY(2024)

引用 0|浏览1
暂无评分
摘要
Goal: Microbubbles (MBs) are known to occur within the circuits of cardiopulmonary bypass (CPB) systems, and higher-order dysfunction after cardiac surgery may be caused by MBs as well as atheroma dispersal associated with cannula insertion. As complete MB elimination is not possible, monitoring MB count rates is critical. We propose an online detection system with a neural network-based model to estimate MB count rate using five parameters: suction flow rate, venous reservoir level, perfusion flow rate, hematocrit level, and blood temperature. Methods: Perfusion experiments were performed using an actual CPB circuit, and MB count rates were measured using the five varying parameters. Results: Bland-Altman analysis indicated a high estimation accuracy (R-2 > 0.95, p < 0.001) with no significant systematic error. In clinical practice, although the inclusion of clinical procedures slightly decreased the estimation accuracy, a high coefficient of determination for 30 clinical cases (R-2 = 0.8576) was achieved between measured and estimated MB count rates. Conclusions: Our results highlight the potential of this system to improve patient outcomes and reduce MB-associated complication risk.
更多
查看译文
关键词
Cardiac surgery,cardiopulmonary bypass,microbubbles,neural network,online detection
AI 理解论文