A conformable phased-array ultrasound patch for bladder volume monitoring

Nature Electronics(2023)

引用 0|浏览17
暂无评分
摘要
Ultrasound can be used to image soft tissues in vivo for the early diagnosis and monitoring of disease progression. However, conventional ultrasound probes are rigid, have a narrow field of view and are operator dependent. Conformable transducers have been proposed, but they lack efficient element localization and effective spatial resolution during mechanical deformations. Here we report a conformable ultrasound bladder patch that is based on multiple phased arrays embedded in a stretchable substrate and can provide mechanically robust, conformable and in vivo volumetric organ monitoring. The phased arrays use Sm/La-doped Pb(Mg 1/3 Nb 2/3 )O 3 –PbTiO 3 ceramics as the piezoelectric material, which offers superior properties ( d 33 = 1,000 pC N −1 , ε r = 7,500 and k 33 = 0.77) than conventional piezoelectric ceramics. We use the conformable ultrasound patch in a pilot clinical study of bladder monitoring. Bladder volume estimation with the patch is comparable (relative errors of 3.2 ± 6.4% and 10.8 ± 8.2% with and without ultrasound gel, respectively) to that obtained using standard clinical ultrasound equipment, and not requiring manual translation or rotation by an operator.