Conformable ultrasound breast patch for deep tissue scanning and imaging

SCIENCE ADVANCES(2023)

引用 2|浏览14
暂无评分
摘要
Ultrasound is widely used for tissue imaging such as breast cancer diagnosis; however, fundamental challenges limit its integration with wearable technologies, namely, imaging over large-area curvilinear organs. We introduced a wearable, conformable ultrasound breast patch (cUSBr-Patch) that enables standardized and reproducible image acquisition over the entire breast with less reliance on operator training and applied transducer compression. A nature-inspired honeycomb-shaped patch combined with a phased array is guided by an easy-to-operate tracker that provides for large-area, deep scanning, and multiangle breast imaging capability. The in vitro studies and clinical trials reveal that the array using a piezoelectric crystal [Yb/Bi-Pb(In1/2Nb1/2)O-3-Pb(Mg1/3Nb2/3)O-3-PbTiO3] (Yb/Bi-PIN-PMN-PT) exhibits a sufficient contrast resolution (similar to 3 dB) and axial/lateral resolutions of 0.25/1.0 mm at 30 mm depth, allowing the observation of small cysts (similar to 0.3 cm) in the breast. This research develops a first-of-its-kind ultrasound technology for breast tissue scanning and imaging that offers a noninvasive method for tracking real-time dynamic changes of soft tissue.