Timing performance of a Micro-Channel-Plate Photomultiplier Tube

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2020)

引用 12|浏览3
暂无评分
摘要
The spatial dependence of the timing performance of the R3809U-50 Micro-Channel-Plate PMT (MCP-PMT) by Hamamatsu was studied in high energy muon beams. Particle position information is provided by a GEM tracker telescope, while timing is measured relative to a second MCP-PMT, identical in construction. In the inner part of the circular active area (radius r<5.5 mm) the time resolution of the two MCP-PMTS combined is better than 10 ps. The signal amplitude decreases in the outer region due to less light reaching the photocathode, resulting in a worse time resolution. The observed radial dependence is in quantitative agreement with a dedicated simulation. With this characterization, the suitability of MCP-PMTS as t(0) reference detectors has been validated.
更多
查看译文
关键词
MCP-PMT,Time resolution,Cherenkov light,Beam test,t0-reference,Monte-Carlo simulation